Repository of Research and Investigative Information

Repository of Research and Investigative Information

Shahid Sadoughi University of Medical Sciences

The optimization of an energy-dispersive X-ray diffraction system for potential clinical application.

(2010) The optimization of an energy-dispersive X-ray diffraction system for potential clinical application. Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine. pp. 2237-45. ISSN 1872-9800

[img] Text
395.pdf

Download (854kB)

Abstract

In the past decade, energy-dispersive X-ray diffraction (EDXRD) has been used to identify the nature of tissues. However, these systems have limited clinical use because of problems such as the long measurement times. In this study, the relation between various setup parameters and some performance specifications such as sensitivity, spatial resolution and momentum transfer resolution were assessed using both geometrical calculations and modeling. Accuracy of the derived relations was also confirmed by means of experimental measurements. As an example, the optimum parameters were determined for obtaining diffraction patterns of breast tissue for an efficient acquisition time. Accordingly, the results of this study could introduce a useful tool for EDXRD optimization in clinical application.

Item Type: Article
Page Range: pp. 2237-45
Journal or Publication Title: Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine
Volume: 68
Number: 12
ISSN: 1872-9800
Depositing User: ms soheila Bazm
URI: http://eprints.ssu.ac.ir/id/eprint/14064

Actions (login required)

View Item View Item