A comparison of patients absorption doses with bone deformity due to the eos imaging and digital radiology

(2017) A comparison of patients absorption doses with bone deformity due to the eos imaging and digital radiology. Archives of Bone and Joint Surgery. pp. 145-148.

[img] Text
308.pdf

Download (354kB)

Official URL: https://www.scopus.com/inward/record.uri?eid=2-s2....

Abstract

Background: This study has aimed to measure the patient dose in entire spine radiography by EOS system in comparison with the digital radiography. Methods: EOS stereo-radiography was used for frontal and lateral view spine imaging in 41 patients in a prospective analytical study. A calibrated dose area product (DAP) meter was used for calibration of the DAP in EOS system. The accuracy and precision of the system was confirmed according to the acceptance testing. The same procedure was used for 18 patients referred for lumbar spine digital radiology (overall 36 images). Results: Although radiation fields in the EOS were almost twice of that in digital radiology, and the average peak tube voltage (kVp), current supply to the tube (mA), and the average size and age of the patients referred for EOS imaging were greater than digital radiology, however, the average DAP in EOS was 1/5 of that in digital radiology system. Also, the average dose in the EOS was about 1/20 of that in digital radiology. Conclusion: The patient dose in EOS imaging system was lower in comparison with digital radiology (1/20). © BY THE ARCHIVES OF BONE AND JOINT SURGERY 2017.

Item Type: Article
Page Range: pp. 145-148
Journal or Publication Title: Archives of Bone and Joint Surgery
Volume: 5
Number: 3
Publisher: Mashhad University of Medical Sciences
Depositing User: ms soheila Bazm
URI: http://eprints.ssu.ac.ir/id/eprint/10025